Glial cell surrounded by axons in a rat neuronal culture labeled for microtubules and actin
Dr. Christophe Leterrier, NeuroCyto, INP, Marseille, France

AX / AX R with NSPARC

Confocal Microscope System

Specifications

 

AX

AX R

Scan Head

FN 25 galvano scanner
Up to 8192 x 8192 pixels
Up to 10 fps at 512 x 512 pixels
Pixel dwell time up to 0.2 microseconds
Supports bidirectional imaging and line-scan imaging

FN 25 resonant scanner
FN 25 galvano scanner
Up to 2048 x 2048 pixels for 2K (1024 x 1024 pixels for 1K)
Up to 720 fps at 2048 x 16 pixels for 2K (720 fps at 1024 x 16 pixels for 1K)
30 fps at 2048 x 512 pixels for 2K (1024 x 512 pixels for 1K)
Supports bidirectional imaging and line-scan imaging

Scan Head input/output port

One FC fiber laser input port and one FC fiber signal output port
An additional FC fiber laser input port and FC fiber signal output port can be optionally added

Laser

Up to 16 wavelengths¹ of visible/near-IR lasers can be installed (when combined with the optional laser input port and AX-LU-EXP AX laser expansion unit)
Compatible laser: (Standard) 405 nm to 750 nm, (option) 730 nm or 785 nm for NIR Imaging Option

Dichroic Mirror

Up to 6 customizable mirrors

DUX-VB Detector

2 or 4 channels
Freely tunable emission bands with ± 1nm accuracy and up to 66 discrete spectral channels
Up to 12 bandpass filters
Multi-alkali PMT or GaAsP PMT options

DUX-ST Detector

2 or 4 channels 
Up to 18 bandpass filters
Multi-alkali PMT or GaAsP PMT options

NSPARC Detector Unit

Lateral resolution 100 nm*1, Axial resolution 300 nm*1
Equipped with SPPC (Single Pixel Photon Counter) array detector
Up to 7 barrier filters can be mounted, including multiple bandpass filters
With galvano scanner: Can be used with X resolution of 64 to 8192 pixels, Y resolution of 2 to 8192 pixels
With resonant scanner: Can be used with X resolution of 256, 512, 1024 and 2048 pixels, Y resolution of 128 to 2048 pixels

Diascopic Detector

Compact PMT detector

Pinhole

Continuously variable

Field Number

25

Z step

Ti2-E: 0.01 μm, 0.02 μm (with encoder control), FN1 stepping motor: 0.05 μm, Ni-E: 0.025 μm

Compatible microscopes

Ti2-E inverted microscope with maximum FOV of 25 mm
Ni-E and FN1 upright microscopes with maximum FOV of 25 mm

Options

Photostimulation (point raster or digital micromirror)
Fluorescence lifetime imaging including fast FLIM
Piezoelectric Z (or XYZ)
Environmental chamber or enclosure
Other modalities such as TIRF, N-STORM

Software

Nikon NIS-Elements C
Optional modules available
Up to 16 bit images (65,536 gray levels with fine integration) with multiple file output options

Control Workstation

Microsoft Windows® 10 64bit/11 Professional with GPU-accelerated graphics card*2

Recommended installation conditions

Temperature 23 ± 5˚C, Humidity 70% RH or less (no condensation)

*1 These values were measured using 40 nm diameter beads for lateral resolution and 100 nm diameter beads for axial resolution with excitation by a 488 nm laser. Actual resolution depends on laser wavelength and optical configuration.

*2 Windows 10 and Windows 11 are trademarks or registered trademarks of Microsoft Corporation in the United States and other countries.