Specifications
CSU-W1 SoRa | CSU‐W1 | CSU‐X1 | |
---|---|---|---|
Confocal Scanning Method | Microlens-enhanced Nipkow disk scanning with additional microlensed emission pinhole SoRa disk | Microlens‐enhanced Nipkow disk scanning | |
Maximum Disk Speed | 4,000 RPM | 5,000 or 10,000 RPM | |
External synchronization | Scan‐speed synchronization through TTL input | ||
Disk Unit | 50um or 25um, and second SoRa super resolution disk, motorized | 50um or 25um (up to 2) motorized | 50um |
Lens-switched light path | 1x, 2.8x, and 4x magnfication | n/a | |
Brightfield | n/a | Optional motorized brightfield bypass | |
Field of View | 61x57um @ 100x in SoRa mode 71x67um@ 60x in SoRa mode |
17x16mm maximum | 10x7mm |
Excitation Wavelengths | 405-640nm | 405nm to 785nm | 405nm to 647nm |
Laser Introduction | Single mode optical fiber | ||
Excitation Shutter | Built‐in mechanical shutter | ||
Dichroic Mirror | 3 position motorized | 3 position motorized or 1 position manual | |
Emission Filter wheel | 10 position | 6 or 12 position | |
External Control | RS‐232C | ||
Microscope Mount | Direct microscope coupler | C‐Mount coupler | |
Camera adapter | C‐mount 1x (variable magnifications available to match sensor sizes) | ||
Operating Environment | 15‐35°C, 20‐75% RH | 15‐40°C, 20‐75% RH | |
Power | Input: 100‐240 VAC +/‐ 10%, 50/60 Hz | ||
Microscope Compatibility | Ti2 series, Ti series | Ti2 series, Ti series, Ni series, FN1 series |