Specifications
ECLIPSE Ji | |
---|---|
Observation methods | Brightfield, Epi-fluorescence |
Optical system | CFI Infinity Optical System Observation Optical System: Inverted image observation FOV 25 Optical path switching: Switching between the built-in camera optical system and the left side port |
Built-in camera | Imaging device: 7.8 megapixel monochrome CMOS sensor Output signal Tone: Monochrome 12 bits/8 bits Frame rate: Maximum 18 fps Output pixel number: 2800×2800 pixels (when assay used) |
Focusing | Drive system: Motorized (Via PFS nosepiece objective lens up/down movement)Focusing stroke: About 10 mm Focusing speed: Maximum driving speed 2.5 mm/sec |
PFS* | Focal point maintenance control: Infrared light projecting method Applicable observation methods: Brightfield, Fluorescence observation |
Transmission illumination section | Koehler illumination Light source: LED |
Stage | Stroke: X: ±59 mm, Y: ±39.5 mm Maximum drive speed About 25 mm/sec |
Nosepiece | Objective lens mounting holes: 6 Nosepiece drive method: Motorized |
Fluorescence cube turret | Number of filter cubes that can be mounted: 6 (Compatible with wide-field filter cubes) Turret drive method: Motorized |
Light distribution section | Light source used: D-LEDI2 fluoresce LED light source |
PC interface | USB interface: Device interface (for built-in camera) B connector USB 3.0 (SuperSpeed) |
Input rating | 100V-240VAC±10%, 3.0 A, 50/60 Hz |
Power consumption | 320 W |
Power source cord | - 100 to 120 V: Power source cord of 3 conductor grounding Type SVT, NO.18 AWG, 3 m long maximum, rated at 125VAC minimum with detachable receptacles conforming to UL specifications - 220 to 240 V: Power source cord of 3 conductor grounding Type H05VV-F 1 mm2, 3 m long maximum, rated at 250VAC minimum with detachable receptacles conforming to EU/EN specifications |
*PFS: a function that automatically corrects focal point displacement due to the passage of time and/or stage movement.