11.
Preparation
Adjusting the condenser
MENU
CLOSE
Preparation
Observe comfortably
Carrying the microscope
Set up
ECO mode and sleep function
Instructions for how to use the Light Intensity Management
Lamp illumination and adjusting the brightness
Adjusting the interpupillary distance
Diopter adjustments (1)
Diopter adjustments (2)
Diopter adjustments (3)
Stage vertical movement stopper
Adjusting the condenser
Observation
Observing
Making observations using light field microscopy
Making observations using phase contrast microscopy
Making observations using transmitted light fluorescence microscopy
Making observations using darkfield microscopy
Storage
After use and storage
After use and storage
Cleaning the eyepieces
Advanced information
Observe at maximum magnification
Oil immersion operation
Cleaning the oil immersion lens
Name
Optical system of microscope
Optical system of microscope
Name of each part
Information for the Nikon website
Top page
Contact
Adjusting the condenser
Focus the field diaphragm image
Tighten the aperture diaphragm until you can see the field diaphragm image in the viewfield.
Rotate the condenser vertical movement knob and focus on the field diaphragm's image until the edges can be seen clearly.
Rotate the condenser centering knob to center the condenser. Move the condenser to center the positioning of the diaphragm image in the field of view.
Rotate the field diaphragm dial and adjust the image's size so that the diaphragm image circumscribes the field of view.
Adjust the aperture diaphragm lever
Set the aperture diaphram lever of the condenser lens to "the same value as the NA of the objective lens".
By doing this, the aperture diaphragm will limit to 70 to 80% of the objective's numerical aperture, allowing sufficient contrast.
PREV
NEXT
View the list of all content