| CSU-W1 SoRa | CSU‐W1 | |
|---|---|---|
| Confocal Scanning Method | Microlens-enhanced Nipkow disk scanning with additional microlensed emission pinhole SoRa disk | Microlens‐enhanced Nipkow disk scanning | 
| Maximum Disk Speed | 4,000 RPM | |
| External synchronization | Scan‐speed synchronization through TTL input | |
| Disk Unit | 50um or 25um, and second SoRa super resolution disk, motorized | 50um or 25um (up to 2) motorized | 
| Lens-switched light path | 1x, 2.8x, and 4x magnfication | n/a | 
| Brightfield | n/a | Optional motorized brightfield bypass | 
| Field of View | 61x57um @ 100x in SoRa mode 71x67um@ 60x in SoRa mode | 17x16mm maximum | 
| Excitation Wavelengths | 405-640nm | 405nm to 785nm | 
| Excitation Shutter | Built‐in mechanical shutter | |
| Dichroic Mirror | 3 position motorized | |
| Emission Filter wheel | 10 position | |
| External Control | RS‐232C | |
| Microscope Mount | Direct microscope coupler | |
| Camera adapter | C‐mount 1x (variable magnifications available to match sensor sizes) | |
| Operating Environment | 15‐35°C, 20‐75% RH | |
| Power | Input: 100‐240 VAC +/‐ 10%, 50/60 Hz | |
| Microscope Compatibility | Ti2 series, Ti series | Ti2 series, Ti series, Ni series, FN1 series |